QCM-D vs. SPR – What are the differences, and which technique fits your application?

When choosing between two real-time, surface sensitive, and label-free technologies such as Quartz crystal microbalance with dissipation monitoring (QCM-D) and Surface plasmon resonance spectroscopy (SPR), it is important to understand what differentiates the two techniques. Both are used in surface interaction analysis to monitor specific interactions, molecular binding, adsorption and desorption events, and degradation. While the techniques are similar, there are also significant differences when it comes to information extraction/output and experimental capabilities. In this overview, we sort out the key differences and list examples of suitable scenarios for using each method.

What are your needs?

Comparing the two instruments, you will see a set of differences. Some of which are inherent in the measurement principle, and some which originate in the hardware design. Therefore, prior to deciding which method to use, you must define what information you need, and what capabilities are necessary for your experiments.

Which technique can deliver the answers you are looking for, and does the supplier of that technology offer all the possible features you need to run the experiments?


Differences and similarities in information output between QCM-D and SPR

Starting with the respective technology and measurement principle, QCM (including the extended versions, QCM-D, QCM-I, QCM-A, etc.) is an acoustic technology, and SPR is an optical one. In brief, QCM measures resonance frequency changes, f, of an oscillating quartz crystal, while SPR measures changes of the surface plasmon resonance angle, ϴ. The resonance frequency of the crystal oscillation is sensitive to mass changes, and the SPR angle is sensitive to refractive index changes.

In addition to f and ϴ , some instruments capture additional parameters as well, expanding the amount of information extracted from each experiment. Depending on which captured parameters are offered by a specific QCM supplier, information output from QCM technology ranges from mere qualitative information about time-resolved mass changes at the surface, to quantitative information about mass, thickness, and viscoelastic properties at high time-resolution. The latter is the case for QSense® QCM-D, which is the QCM-setup that we will be referring to throughout this post.

Information output offered by a standard SPR instrument ranges from relative information about refractive index changes at/near the surface, to quantitative information about the refractive index, thickness, density, and surface coverage. Below we give examples of typical differences between the two technologies.

Examples of differences between QCM-D and SPR data output


Experimental capabilities and hardware limitations

Just because measurements are possible in principle does not mean hardware is always set up to support them in practice. Instrument suppliers choose to package the technologies in different ways, optimizing for usability, footprint or other aspects. The experimental capabilities and potential limitations of the hardware design are, therefore, important to assess when deciding which experimental method to use. For example, are you able to mimic the relevant conditions such as surface material, solvent, and temperature? Can you afford the amount of sample required to run the set of experiments? The hardware design can both limit experimental possibilities, as well as open new opportunities.


Measurement principle and hardware design

As already discussed, some of the differences between the two technologies are inherent in the measurement principle, while other restrictions stem from the hardware design. In table 1, we list a side-by-side comparison of typical information output, and hardware specifications to give a general idea of what can be expected from a QSense QCM-D and a standard SPR. Please note that the respective listed specifications may differ between instrument suppliers.

QSense QCM-DSPR
TechnologyAcoustic techniqueOptical technique
Measured parametersShifts of frequency, f, and dissipation, D, at multiple harmonics of an oscillating quartz crystalShifts of plasmon resonance angle, ϴ 1


Information outputQCM-D monitors time-resolved relative mass changes and layer softness, which provide info on:
• Binding/adsorption/desorption/degradation
• Uptake/removal rates
• Conformational changes (swelling/crosslinking/collapse)

When the model assumptions are fulfilled, it is possible to quantify absolute changes of:
• Acoustic mass (incl. coupled solvent)
• Thickness
• Viscoelastic properties

Traditional SPR monitors time-resolved, relative refractive index changes (related to relative mass changes), which provides info on:
• Binding/adsorption/desorption/degradation
• Kinetic rate constants
• Affinity constants
• Concentration

In certain situations, it can be possible to estimate or quantify [5]:
• Thickness
• Refractive index
• Optical mass (i.e. molar mass)
Substrate material• Versatile
• In general, any coating that is thin and rigid can be used as a substrate material




• Noble metal is needed to excite the plasmons. Most SPRs use gold.
• Thin coatings can be put on top of metal, but thick coatings or coatings with high refractive index will generally quench the signal as the evanescent field decays exponentially in the metal and dielectric media [1]
MediumGas phase and liquid


The technology as such works in both gas phase and liquid, but some suppliers have restricted the usage to only liquid
Dynamic thickness rangeTypically, 10-10 m to 10-6 m

The maximum thickness is limited by the penetration depth of the shear wave into the medium.
Typically, < 10-10 m to 2-3 10-7 m

The maximum thickness is limited by the exponential decay of the evanescent field.

Senses water contentYesNo
Senses structural changesCan sense structural changes of molecular layers, such as swelling, crosslinking, and collapseTraditional SPR cannot sense structural changes. Some SPRs can sense swelling [3, 6]
Characterization of material propertiesYesNo
KineticsCan measure kinetic values however fluid exchange is not as rapid, and mass changes include associated solventA standard for measuring kinetic constants and affinity

Sample volume over surface15 – 40 µlTypically, < µl – ~ µl [2, 3]
Min. sample volume per measurement channel~ 50 µl – 300 µl per channel2 – 5000 µl per channel [2, 3, 4]
Table 1: Overview of the main technology and hardware differences between QSense QCM-D and SPR, based on specifications listed at different suppliers’ websites [2-4]

Which technology is right for my application?

Depending on what questions you are asking, and what your experimental needs are, one technology may be more suitable than the other. In Table 2, we list a few questions and experimental needs to guide you to which technology is the most appropriate for your application.

If you would like to…QSense QCM-DSPR
Study binding
Study adsorption and desorption
Study degradation
Characterize hydrated mass
Characterize non-hydrated mass, i.e. molar mass(√)
Characterize layer thickness(√)
Characterize material properties (viscoelasticity)
Characterize conformational changes such as swelling, crosslinking, or collapse (i.e. molar mass unchanged)(√)
Quantify the number of molecules adsorbed/bound
Extract binding constants
Extract affinity
Quantify concentration
Use various substrate materials such as e.g. plastics, glass, or custom coatings
Use harsh solvents(√)
Work at high temperatures(√)
Limit sample consumption to maximum ~µl(√)
Table 2: Suitability of the respective technology to address certain scientific questions, where √ marks suitable, (√) means that the suitability depends on the specific conditions or the specific instrument used, and no checkmark indicates less/not suitable.

Concluding remarks

QCM-D and SPR are two surface sensitive real-time technologies with overlapping application areas and which, at first glance, may appear seemingly alike. Similarities do exist, but there are also differences that may be critical depending on the scientific question asked, or the experimental needs. Overall, the acoustic QCM-D and optical SPR are complementary, but there are applications for which one or the other may be more suitable. In this overview, we have presented the key differences and the circumstances under which each technology is most suitable.


References

  1. S. Ekagasit, et. al., Applied Spectroscopy, 59, 5, 2005, p 661 – 667
  2. http://web.bf.uni-lj.si/bi/sprcenter/BiacoreT100.pdf
  3. http://www.bionavis.com/en/life-science/technology/traditional-spr/
  4. http://www.reichertspr.com/index.php?src=directory&view=systems&srctype=systems_detail_comparison-chart&refno=352
  5. Reimhult et. al., Anal. Chem, 76 (24) 2004, 7211-7220
  6. https://www.sprpages.nl/spr-over-view/spr-theory#ref4

Footnotes

  1. This is the parameter most commonly measured. There are, however, other ways to measure as well, listed in ref [6]
Featured News
Contact Nanoscience
Request More info

Want to learn more? Talk to a scientist:

Related News & Posts

More from Nanoscience

The Role of Scanning Electron Microscopy in the Automotive Industry

From zippy city scooters to heavy duty haulers, modern automobiles are assemblies of thousands of precisely manufactured comp…

September 20, 2023

7 Factors to Consider While Buying an Electrospinning Machine

Electrospinning is a versatile nanofiber fabrication technique wherein a polymer solution is charged and ejected from a spinn…

September 13, 2023

Applications of SEM in Semiconductor Device Failure Analysis

Failure analysis (FA) is an essential methodology that engineers use in the development process of semiconductor devices, lik…

September 6, 2023

This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.